A simple method was developed to determine 10 organic acids simultaneously in tomato products using reverse-phase high performance liquid chromatography (HPLC) column with the diode array detector set at 210 nm. After centrifugation and filtration, the samples were passed through to an anion exchange resin and the organic acids were released using 0.1 N HCl. The chromatographic separation was achieved with isocratic analysis in a 20-min run. The method was reliable and sensitive. The coefficient of determination of the standard calibration curve is 0.9925 r2 0.9999 and the limit of detection ranged from 0.08 to 6.00 mg/kg for trans-aconitic acid and acetic acid, respectively. The limit of quantification ranged from 0.19 to 15.18 mg/kg for trans-aconitic and acetic acid, respectively. To establish the efficiency of the anion resin the procedure was applied to a standard solution of a mixture of organic acids. The organic acids recovery ranged from 87.0% 1.9 for citramalic acid to 109.9% 5.2 for fumaric acid.

Organic acids profile in tomato juice by HPLC with UV detection / Marconi, O.; Floridi, S.; Montanari, Luigi. - In: JOURNAL OF FOOD QUALITY. - ISSN 0146-9428. - 30:(2007), pp. 43-56.

Organic acids profile in tomato juice by HPLC with UV detection

MONTANARI, Luigi
2007-01-01

Abstract

A simple method was developed to determine 10 organic acids simultaneously in tomato products using reverse-phase high performance liquid chromatography (HPLC) column with the diode array detector set at 210 nm. After centrifugation and filtration, the samples were passed through to an anion exchange resin and the organic acids were released using 0.1 N HCl. The chromatographic separation was achieved with isocratic analysis in a 20-min run. The method was reliable and sensitive. The coefficient of determination of the standard calibration curve is 0.9925 r2 0.9999 and the limit of detection ranged from 0.08 to 6.00 mg/kg for trans-aconitic acid and acetic acid, respectively. The limit of quantification ranged from 0.19 to 15.18 mg/kg for trans-aconitic and acetic acid, respectively. To establish the efficiency of the anion resin the procedure was applied to a standard solution of a mixture of organic acids. The organic acids recovery ranged from 87.0% 1.9 for citramalic acid to 109.9% 5.2 for fumaric acid.
2007
Organic acids profile in tomato juice by HPLC with UV detection / Marconi, O.; Floridi, S.; Montanari, Luigi. - In: JOURNAL OF FOOD QUALITY. - ISSN 0146-9428. - 30:(2007), pp. 43-56.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11388/61267
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 20
  • ???jsp.display-item.citation.isi??? 10
social impact