A new capillary electrophoresis method to measure human blood plasma arginine and citrulline levels in a single run without derivatization was established. After adding homoarginine as internal standard, plasma proteins were removed by a 90:10 v/v acetonitrile/ammonia mixture. Arginine and citrulline were detected by an ultraviolet detector at 190 nm and separated in 11.65 and 20.43 min, respectively, by using a 75 mmol/L Tris phosphate solution at pH 1.2 as a background electrolyte. Limits of detection were 0.8 and 5 mol/L for arginine and citrulline, respectively. Precision tests indicated a good repeatability of migration times and of peak area both for citrulline (CV% = 0.82 and 3.19) and arginine (CV% = 0.65 and 2.79). The CV% for intra- and interassay tests were, respectively, 1.84 and 3.23 for citrulline and 1.25 and 1.50 for arginine. Mean recovery was 101.5 and 98.5% for citrulline and arginine, respectively. The performance of the developed method was assessed by measuring plasma arginine levels in 52 subjects and the data were compared with those obtained by our previous assay. The new method was then applied to assess plasma citrulline and arginine in ten chronic kidney disease patients under hypolipidemic therapy with statin.

Simultaneous determination of citrulline and arginine in human blood plasma by capillary electrophoresis with ultraviolet absorption detection / Forteschi, M; Sotgia, S; Pintus, G; Zinellu, A; Carru, C.. - In: JOURNAL OF SEPARATION SCIENCE. - ISSN 1615-9314. - 37:(2014), pp. 2418-2423. [doi: 10.1002/jssc.201400177]

Simultaneous determination of citrulline and arginine in human blood plasma by capillary electrophoresis with ultraviolet absorption detection

Sotgia S;Carru C.
2014-01-01

Abstract

A new capillary electrophoresis method to measure human blood plasma arginine and citrulline levels in a single run without derivatization was established. After adding homoarginine as internal standard, plasma proteins were removed by a 90:10 v/v acetonitrile/ammonia mixture. Arginine and citrulline were detected by an ultraviolet detector at 190 nm and separated in 11.65 and 20.43 min, respectively, by using a 75 mmol/L Tris phosphate solution at pH 1.2 as a background electrolyte. Limits of detection were 0.8 and 5 mol/L for arginine and citrulline, respectively. Precision tests indicated a good repeatability of migration times and of peak area both for citrulline (CV% = 0.82 and 3.19) and arginine (CV% = 0.65 and 2.79). The CV% for intra- and interassay tests were, respectively, 1.84 and 3.23 for citrulline and 1.25 and 1.50 for arginine. Mean recovery was 101.5 and 98.5% for citrulline and arginine, respectively. The performance of the developed method was assessed by measuring plasma arginine levels in 52 subjects and the data were compared with those obtained by our previous assay. The new method was then applied to assess plasma citrulline and arginine in ten chronic kidney disease patients under hypolipidemic therapy with statin.
2014
Simultaneous determination of citrulline and arginine in human blood plasma by capillary electrophoresis with ultraviolet absorption detection / Forteschi, M; Sotgia, S; Pintus, G; Zinellu, A; Carru, C.. - In: JOURNAL OF SEPARATION SCIENCE. - ISSN 1615-9314. - 37:(2014), pp. 2418-2423. [doi: 10.1002/jssc.201400177]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11388/80696
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 13
  • ???jsp.display-item.citation.isi??? 12
social impact