Purpose: Planar cardiac scintigraphy with [123I]I-mIBG is widely used to image myocardial sympathetic activity. In multicentre studies, different γ-cameras can cause variability in quantitative parameters. This study aimed to harmonise multicentre [123I]I-mIBG data using a custom-designed phantom and to assess its feasibility and acceptability. Methods: A ‘tube and bottle’ phantom was designed to standardise the heart-to-mediastinum (H/M) ratio across 15 centres. Each centre prepared three versions of the phantom (A, B, and C) with varying pertechnetate ([99mTc]Tc-O4-) activities, acquired static images using their own γ-camera, and uploaded DICOM data to a shared platform. In the phantom, the tube and bottle represent the heart and mediastinum, respectively, with the tube-to-bottle (T/B) ratio simulating the H/M ratio. The reference centre analysed the images and calculated the T/B ratios, applying linear regression for data harmonisation. A survey was conducted to assess the phantom’s usability. Results: The harmonised T/B ratios for versions A and B were 20.46 ± 0.78 and 6.19 ± 0.39, respectively. The average slope and intercept of the regression line across the participating centres resulted in 1.07 ± 0.38 and − 0.82 ± 4.95. Survey feedback indicated high feasibility and acceptability, with all centres recommending the phantom for multicentre harmonisation. Conclusions: The custom-made phantom effectively harmonised experimental data across different centres, supporting its use in multicentre studies to improve data consistency.

Multi-centre data harmonisation applied to heart-to-mediastinum quantification in parkinsonism (ITA-MIBG): a cross-calibration phantom study with tube and bottle / Filippi, Luca; Bianconi, Francesco; Minestrini, Matteo; De Vincentis, Giuseppe; Frantellizzi, Viviana; Marongiu, Andrea; Ferrari, Cristina; Giordano, Alessia; D'Ambrosio, Daniela; Gazzilli, Maria; Bertagna, Francesco; Bagni, Oreste; Panareo, Stefano; Guglielmo, Priscilla; Lazzarato, Achille; Casali, Massimiliano; Aghakhanyan, Gayane; Spanu, Angela; Evangelista, Laura; Schillaci, Orazio; Palumbo, Barbara. - In: CLINICAL AND TRANSLATIONAL IMAGING. - ISSN 2281-7565. - (2025). [10.1007/s40336-025-00681-4]

Multi-centre data harmonisation applied to heart-to-mediastinum quantification in parkinsonism (ITA-MIBG): a cross-calibration phantom study with tube and bottle

Marongiu, Andrea;Lazzarato, Achille;Spanu, Angela;Palumbo, Barbara
2025-01-01

Abstract

Purpose: Planar cardiac scintigraphy with [123I]I-mIBG is widely used to image myocardial sympathetic activity. In multicentre studies, different γ-cameras can cause variability in quantitative parameters. This study aimed to harmonise multicentre [123I]I-mIBG data using a custom-designed phantom and to assess its feasibility and acceptability. Methods: A ‘tube and bottle’ phantom was designed to standardise the heart-to-mediastinum (H/M) ratio across 15 centres. Each centre prepared three versions of the phantom (A, B, and C) with varying pertechnetate ([99mTc]Tc-O4-) activities, acquired static images using their own γ-camera, and uploaded DICOM data to a shared platform. In the phantom, the tube and bottle represent the heart and mediastinum, respectively, with the tube-to-bottle (T/B) ratio simulating the H/M ratio. The reference centre analysed the images and calculated the T/B ratios, applying linear regression for data harmonisation. A survey was conducted to assess the phantom’s usability. Results: The harmonised T/B ratios for versions A and B were 20.46 ± 0.78 and 6.19 ± 0.39, respectively. The average slope and intercept of the regression line across the participating centres resulted in 1.07 ± 0.38 and − 0.82 ± 4.95. Survey feedback indicated high feasibility and acceptability, with all centres recommending the phantom for multicentre harmonisation. Conclusions: The custom-made phantom effectively harmonised experimental data across different centres, supporting its use in multicentre studies to improve data consistency.
2025
Multi-centre data harmonisation applied to heart-to-mediastinum quantification in parkinsonism (ITA-MIBG): a cross-calibration phantom study with tube and bottle / Filippi, Luca; Bianconi, Francesco; Minestrini, Matteo; De Vincentis, Giuseppe; Frantellizzi, Viviana; Marongiu, Andrea; Ferrari, Cristina; Giordano, Alessia; D'Ambrosio, Daniela; Gazzilli, Maria; Bertagna, Francesco; Bagni, Oreste; Panareo, Stefano; Guglielmo, Priscilla; Lazzarato, Achille; Casali, Massimiliano; Aghakhanyan, Gayane; Spanu, Angela; Evangelista, Laura; Schillaci, Orazio; Palumbo, Barbara. - In: CLINICAL AND TRANSLATIONAL IMAGING. - ISSN 2281-7565. - (2025). [10.1007/s40336-025-00681-4]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11388/358489
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 1
social impact